Patient Monitor User Manual Monitoring CO
2
- 155 -
As long as no sampling line is connected, the Masimo Sidestream CO
2
module remains in a
low-power sleep mode. Once the sampling line is connected, the Masimo Sidestream CO
2
module switches to measuring mode and starts delivering gas data.
For ordering information about NomoLine sampling lines, cannulas, and related consumables,
visit www.masimo.com.
Sampling Line Replacement
NomoLine sampling lines should be replaced between each patient or when the sampling line
becomes occluded. Occlusion occurs when water, secretions etc. are aspirated from the
respiratory circuit to such an extent that NomoLine capnography cannot maintain the normal 50
ml/min sample flow. This is indicated by a red flashing LEGI indicator and an alarm message;
replace the sampling line and wait until the LEGI indicator switches to green, indicating that the
Masimo Sidestream CO
2
module is again ready for use.
For ordering information about NomoLine sampling lines and related consumables, visit
www.masimo.com.
Setting up
1. Remove the temporary plug from the NomoLine Capnography input connector.
2. Securely mount the Masimo sidestream CO
2
module.
Note: A bracket designed to mount the module is available, visit www.masimo.com.
3. Connect the gas sample exhaust port on the rear of the module to a scavenging system if
intended to be used in combination with N
2
O and/or anesthetic agents.
4. Connect the module cable to the connection port of the monitor.
5. Connect a NomoLine sampling line to the module input connector.
6. Check that the monitor is powered up and correctly configured.
7. Set
Work Mode
to
Measure
.
8. Check that the gas inlet indicator shows a steady green light, indicating that the Masimo
sidestream CO
2
module is ready for use.
Note: Without a sampling line connected, the gas inlet indicator does not illuminate.
9. Attach a NomoLine sampling line to the patient for monitoring. Refer to the NomoLine
sampling line Directions for Use.
10. Following connection of the NomoLine sampling line, check that CO
2
values appear on the
monitor screen.
Operation
The information in this section assumes that Masimo sidestream CO
2
module is set up and ready
for use. This section provides necessary information for proper operation of the device. Do not
operate Masimo sidestream CO
2
module without completely reading and understanding these
instructions.
The Light Emitting Gas Inlet (LEGI) Indicator provides visual indications of capnography status.