Patient Monitor User Manual Alarm Information
- 61 -
Message
Cause
Alarm Level
Action Taken
NIBP Pulse Signal
Weak
(Omron NIBP module)
Pulse is too weak, and the
detected signal is too weak.
Low
Check if the cuff
leak or cuff is
properly wrapped.
TEMP
TEMP T1 Sensor Off
Temperature cable of TEMP
channe1 may be
disconnected from the
monitor.
Low
Make sure that the
cable is properly
connected
TEMP T2 Sensor Off
Temperature cable of TEMP
channe2 may be
disconnected from the
monitor.
Low
Make sure that the
cable is properly
connected.
Excessive T1
TEMP1 measuring value is
beyond measuring range.
High
Check sensor
connection and
patient condition
Excessive T2
TEMP2 measuring value is
beyond measuring range.
High
Check sensor
connection and
patient condition
TEMP Comm Fail
TEMP module failure or
communication failure.
High
Stop measuring
function of TEMP
module, and notify
biomedical engineer
or Manufacturer’s
service staff.
T1 Calibration Failed
T1 calibration failed.
High
Please check whether
the module works
properly.
T2 Calibration Failed
T2 calibration failed
High
Please check whether
the module works
properly.
IBP
YY
Sensor Off
(YY
stands for the IBP label
name:
Art
,
PA
,
CVP
,
RAP
,
LAP
,
ICP
,
P1
and
P2
)
IBP sensor falls off.
Medium
Check the sensor
connection and
reconnect the sensor.