Patient Monitor User Manual Monitoring AG
- 175 -
MAC
standard2
is the MAC
standard
value of the secondary anesthetic gas.
MAC
standardN2O
is the MAC
standard
value of N
2
O.
MAC
standard
value is as following:
MAC
standard
HAL 0.77%
ENF 1.7%
ISO 1.15%
DES 6.0%
SEV 2.1%
N
2
O 105%
19.3.2 Monitoring Steps for Masimo ISA Module
NomoLine Sampling Lines
Masimo Sidestream CO
2
module samples gas from the respiratory circuit through the NomoLine
Family sampling line at a rate of 50 sml/min, making measurements of CO
2
possible for adult,
pediatric, infant and neonatal patients. The NomoLine Family of sampling lines are designed for
optimal performance and measurement fidelity when used with the Masimo Sidestream CO
2
module.
NomoLine sampling lines include nasal and nasal/oral cannulas for non-intubated patients with
and without supplementary oxygen delivery and airway adapter sets for intubated patients.
As long as no sampling line is connected, the Masimo Sidestream CO
2
module remains in a
low-power sleep mode. Once the sampling line is connected, the Masimo Sidestream CO
2
module switches to measuring mode and starts delivering gas data.
For ordering information about NomoLine sampling lines, cannulas, and related consumables,
visit www.masimo.com.
Working Status of ISA analyzer
Working status of the ISA analyzer can be indicated by the indicator. For the detailed information,
please refer to the following table.
Indication
Status
Steady green light
System OK
Blinking green light Zeroing in progress