Patient Monitor User Manual Product Specification
- 257 -
Measure Parameter
T1, T2, TD(the absolute value of T2 minus T1)
Channel 2
Sensor Type
YSI-10K and YSI-2.252K
Unit
°
C,
°
F
Measuring Range
0
°
C to 50
°
C(32 ºF to 122 ºF)
Resolution
0.1
°
C (0.1 ºF)
☆
Accuracy
1
±
0.3
°
C
Refresh Time
Every 1 s to 2 s
Temperature Calibration
At an interval of 5 to 10 minutes
Measuring Mode
Direct Mode
Transient Response Time
≤
30 s
Note 1: The accuracy consists of two parts, as following:
Accuracy (not including sensor): ± 0.1
°
C
Sensor accuracy:
≤
±
0.2
°
C
A.10 Quick TEMP
Only applicable to iM50 and M50.
Measuring Range
25
°
C ~ 45
°
C(monitoring mode)
35.5
°
C
~
42
°
C (prediction mode)
Operating Temp
10
°
C ~ 40
°
C
Sensor Type
Oral/Axillary sensor, Rectal sensor
Resolution
0.1
°
C
Accuracy
1
±
0.3
°
C
Response time
< 60 s
Update time
1 s ~ 2 s
Warm-up time
Less than 10 seconds
Prediction time
Less than 30 seconds
Measuring Mode
Direct Mode/Adjusted Mode
Transient Response Time
≤
30 s
Clinical Bias
(-0.2 to -0.4 )
°
C
Limits of Agreement
0.49