Patient Monitor User Manual Networked Monitoring
- 38 -
is in the
Wi-Fi Setup
window, switch the
Network Type
from
Wi-Fi
to Wired and then
to
Wi-Fi
again. Then retry to connect to a wireless network. If the wireless network still
fails to be connected, please try to restart the monitor and connect again.
4
Use the wireless device recommended by EDAN, otherwise some exceptional
situations such as frequent network disconnection may occur on the monitor.
5
The wireless driver is compatible with channels 1-11 only.
6
When signal intensity is level 2 or less, signal may be unstable and quality of the
signal transmission may be degraded.
7
When the monitor is connected to MFM-CMS/Gateway via the wireless network, the
user should set the router to a secure encryption/authentication and use the
non-dictionary password.
Recommended options: WPA/WPA2 Personal (supports AES/TKIP);
Other options: none or WPA/WPA2 Enterprise (includes TLS/TTLS /PEAP).
8
The storage path for certificate related to EAP-TLS enterprise-level ecryption method:
root directory of USB flash drive\certs\wlan\.
WARNING
1 Before monitoring patient,
Network Type
(
Wired
or
Wi-Fi
) should be selected and is
not allowed to switch during monitoring. Otherwise, Wi-Fi may be unavailable.
2 If Wi-Fi is unavailable, please restart the monitor to restore Wi-Fi function under the
precondition of ensuring patient’s safety.
4.3 Network Disconnected Alarms
To configure the network disconnected alarms, select
Menu
>
Maintenance
>
User
Maintain
>
Alarm Setup
and choose
Disconnect Alarm
which can be set to
On
or
Off
. The alarm is off by
default.
NOTE:
1. When the monitor is connected with the central monitoring system, you must set
Disconnect Alarm
to
On
.
2. If
Disconnect Alarm
occurs during audio alarm paused or audio alarm off status, the
monitor will prompt a sounding alarm with information of
NetWork Disconnect
.
During the network disconnected status, activating audio alarm paused or audio alarm
off function can disable the audio alarm signal of
Disconnect Alarm
.
4.4 Connecting the Monitor to MFM-CMS
The monitor can be connected to the central monitoring system (V2.65 and above). Through the
network:
1.
The monitor sends patient information, real-time monitoring or measurement data to the
central monitoring system.