Patient Monitor User Manual Alarm Information
- 73 -
Message
Cause
Zeroing and Calibration
Failed
Zeroing is not performed prior calibration.
Sensor Off,Fail.
Perform calibration when the sensor is off.
Invalid Time,Calibration Fail.
Time is not set up prior calibration.
Unable to Calibrate in Demo
Mode
Perform calibration in Demo Mode.
Calibrating...
Calibration is in progress.
CO
2
Self-Testing
CO
2
module is performing a power-on self-test.
IBP alias collision
The same IBP label appears.
C.O. Lack Param
Parameter is not configured for C.O. measurement.
AG Self-Testing......
(Masimo AG module)
AG module is performing a power-on self-test.
AG Span Calib In Progress
(Masimo AG module)
The calibration of AG module is in progress.
MultiGas Zero in Progress
The zeroing of AG module is in progress.
AG Is Starting
(Dräger AG module)
AG module is starting.
AG Standby
(Dräger AG module)
User sets
Work Mode
to
Standby
.
AG Zero In Progress
(Dräger AG module)
The zeroing of Scio module is in progress.
AG Is Warming Up
(Dräger AG module)
Scio module is warming up and is operating at reduced
accuracy.
AG Changing to Standby
(Dräger AG module)
Work Mode
is switching to
Standby
from
Measure
.
AG Changing to Meas.
(Dräger AG module)
Work Mode
is switching to
Measure
from
Standby
.
AG Agent Low Concentration
(Dräger AG module)
Measured agent concentration is low.
AG Agent Overflow
(Dräger AG module)
The gas concentration has increased above the maximum
threshold.
AG Agent Calculate
(Dräger AG module)
Usually it comes up if no single agent history is available
and a mixture situation occurs.
Содержание iM50
Страница 1: ......
Страница 29: ...Patient Monitor User Manual Basic Operation 17 iM60 iM70 1 2 3 4 5 6 7 8 9 10 11 1 2 3 11 4 5 6 7 8 9 10...
Страница 317: ......