Patient Monitor User Manual Networked Monitoring
- 40 -
4.7 HL7 Communication
The monitor supports HL7 protocol to upload data. Select
Menu
>
User Maintain
and input user
maintain password >
Security
. In this menu:
•
Set
HL7
to
On
/
Off
. To avoid hacker attacking, setting HL7 to
Off
is normally
recommended.
•
Set
HL7 Encryption
method to
Off
or
TLS
(default).
To set HL7 IP address of the client-side, select
User Maintain
>
Network Maintain
.
Click
Import Certificate
to install/upgrade the
Certificate
via USB flash drive. The certificate
issued by Certificate Authority (CA) is recommended and self-signed certificate should be
avoided. For detailed steps of importing certificates, please refers to service manual.
For more information about HL7 communication, refer to
HL7 Communication Protocol Service
Manual
.