Patient Monitor User Manual Product Specification
- 277 -
Gas Sampling Rate
200 mL/min ±20 mL/min
Maximum time until water trap
requires draining
41 hrs (sample gas under BTPS conditions, ambient air
23 °C)
Total System Response Time
< 3 s
Drift Compensation (zeroing)
Automated cyclical zeroing,
once per day (in error-free operation)
Zeroing Duration
< 20 s
Cross Sensitivity
None concerning alcohol (< 3000 ppm blood conc.),
acetone (< 1000 ppm), methane, water vapor, NO, and CO
☆
O
2
☆
Range
(0 to 100) Vol%
☆
Accuracy
1
±
(2.5 Vol% + 2.5 % rel.)
Rise Time (t10…90)
4
< 500 ms
Time to Specified Accuracy
3
< 450 s
☆
CO
2
☆
Range
(0 to 13.6) Vol%
☆
Accuracy
1
±
(0.43 Vol% + 8% rel.)
Rise Time (t10…90)
4
< 350 ms
Time to availability
2
< 60 s
Time to Specified Accuracy
3
< 450 s
☆
N
2
O
☆
Range
(0 to 100) Vol%
☆
Accuracy
1
±
(2 Vol% + 8% rel.)
Rise Time (t10…90)
4
< 350 ms
Time to Specified Accuracy
3
< 450 s
☆
Anesthetic Gases Range
☆
Halothane
(0 to 8.5) Vol%
☆
Isoflurane
(0 to 8.5) Vol%
☆
Enflurane
(0 to 10) Vol%
☆
Sevoflurane
(0 to 10) Vol%
☆
Desflurane
(0 to 20) Vol%