![EDAN iM50 Скачать руководство пользователя страница 296](http://html1.mh-extra.com/html/edan/im50/im50_user-manual_2357173296.webp)
Patient Monitor User Manual Product Specification
- 284 -
A.15.2
Defibrillator Synchronization
Output Impedance
< 500
Ω
Maximum Time Delay
35 ms (R-wave peak to leading edge of pulse)
Waveform Rectangular
wave
Amplitude
High level: 3.5 V to 5.5 V, providing a maximum of 1 mA output
current;
Low level: < 0.5 V, receiving a maximum of 5 mA input current
Minimum Required
R-wave Amplitude
0.3 mV
Pulse Width
100 ms±10%
Limited Current
15 mA rating
Rising and Falling Time
< 1 ms
Interface Type
PS2 connector
A.15.3
Nurse Call
Drive Mode
Voltage output
Power Supply
≤
12.6 VDC, 200 mA Max.
Interface Signal
12 V power supply and PWM waveform
Interface Type
PS2 connector
PS2 connector Definition for Analog Output/Defibrillator Synchronization/Nurse Call
PIN.NO. Signal name
Signal Description
1
ANALOG_OUT
Analog out signal
2 GND
Ground
3
SYS_OUT
Defibrillator Synchronization signal
4
+12V
Nurse call power
5 GND
Ground
6
NURSE_OUT
Nurse call control signal
A 15.4 USB Interfaces
Number of USB Interfaces
Standard:2
Drive Mode
HOST interface, USB1.0/2.0 protocol