Patient Monitor User Manual Product Specification
- 269 -
A.13 C.O.
Not applicable to iM50 or M50.
Technique Thermodilution
Technique
Measure Parameters
C.O., TB, TI
Measuring Range
C.O.
0.1 L/min to 20 L/min
TB
23
°
C to 43
°
C(73.4 ºF to 109.4 ºF)
TI
-1
°
C to 27
°
C(30.2 ºF to 80.6 ºF)
Resolution
C.O. 0.1
L/min
TB, TI
0.1
°
C (+0.1 ºF)
Accuracy
C.O.
±
5% or
±
0.2 L/min, whichever is greater
TB
±
0.1
°
C (not including sensor)
TI
±
0.1
°
C (not including sensor)
NOTE:
At least 90% of the C.O. data should reside inside the bounded region, and the lower
95% confidence interval should not exceed 85%.
A.14 AG
Complies with ISO 80601-2-55: 2011.
EDAN G7 module and Masimo module are applicable to iM70, iM80 and M80. Dräger
Minimodule is only applicable to iM80 and M80.
A.14.1 Sidestream
EDAN G7 Module
Intended
Patient
Adult, pediatric, neonatal
Measure
Parameters
Halothane (HAL), Isoflurane (ISO), Enflurane (ENF), Sevoflurane (SEV),
Desflurane (DES), CO
2
, O
2
, N
2
O, AwRR, and MAC
Unit
HAL, ISO, ENF, SEV, DES, N
2
O: %;
CO
2
, O
2
: mmHg, %, kPa, default is %;
AwRR: bpm;
☆
Measuring CO
2
0~15
vol%