Patient Monitor User Manual EMC Information
- 286 -
B EMC Information
- Guidance and Manufacture’s Declaration
B.1 Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
iM50, iM60, iM70, iM80, M50 and M80 are intended for use in the electromagnetic
environment specified below. The customer or the user of iM50, iM60, iM70, iM80, M50 and
M80 should assure that they are used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
iM50, iM60, iM70, iM80, M50 and M80 use RF
energy only for their internal function. Therefore,
their RF emissions are very low and are not
likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class A
iM50, iM60, iM70, iM80, M50 and M80 are
suitable for use in all establishments, other than
domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
Complies
NOTE:
The EMISSIONS characteristics of iM50/iM60/iM70/iM80/M50/M80 make it suitable for
use in industrial areas and hospitals (CISPR 11 class A). If it is used in a residential
environment (for which CISPR 11 class B is normally required)
iM50/iM60/iM70/iM80/M50/M80 might not offer adequate protection to radio-frequency
communication services. The user might need to take mitigation measures, such as
relocating or re-orienting the equipment.
B.2 Electromagnetic Immunity
Guidance and manufacture’s declaration – electromagnetic immunity
iM50, iM60, iM70, iM80, M50 and M80 are intended for use in the electromagnetic
environment specified below. The customer or the user of iM50, iM60, iM70, iM80, M50 and
M80 should assure that they are used in such an environment.