218
53220A/53230A User’s Guide
6
53220A/53230A Math, Graphing, and Data Logging
•
When ‘Statistics’ (under the Math key) are enabled, the minimum and
maximum values from the reading statistics are used.
•
If readings are in progress and ‘Statistics’ are off, the instrument will select
a minimum and a maximum value from (up to) the last 10,000 readings
taken.
•
If no previous readings are available, a minimum bin range of 0 and a
maximum bin range of 1 000 000 000.0 are set.
Automatic Range Setting
Alternately, lower and upper range values can be set automatically based on a
minimum value and maximum value obtained from the first ‘n’ number of
readings taken.
Automatic range selection is enabled with the commands:
CALCulate2:TRANsform:HISTogram:RANGe:AUTO {OFF|ON}
CALCulate2:TRANsform:HISTogram:RANGe:AUTO?
(query form)
ON
- enables automatic selection of the histogram lower and upper range val-
ues.
OFF
- disables automatic selection.
Following a reset (
*RST
) or instrument preset (
SYSTem:PRESet
) automatic
range selection is
enabled
.
The number of readings from which the minimum and maximum range values
are selected (
ON
) is set with the command:
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...