53220A/53230A User’s Guide
169
Triggering and Gating
5
Continuous Totalizing
Continuous totalizing as configured by the command:
CONFigure:TOTalize:CONTinuous
sets the gate source to
TIME
and the gate time to
INFinity
. The command
also sets the input threshold to 0.0V and the edge (events that are totalized) to
positive. Refer to Chapter 4 - “53220A/53230A Input Signal Conditioning” for
information on the commands in the
INPut
subsystem used to change these
parameters.
Reading the Current Count
During continuous or timed totalizing with long
gate times, the current count can be read with the command:
[SENSe:]TOTalize:DATA?
For example:
CONF:TOT:CONT
// configure continuous totalizing
INIT
// start measurement
.
SENS:TOT:DATA?
// query current count
.
ABOR
// terminate measurements