80
53220A/53230A User’s Guide
3
53220A/53230A Measurements
Using
CONFigure
and the appropriate low-level commands, the configuration
can be changed before the measurement is initiated (the abbreviated form of
the SCPI commands are shown):
//configure counter for frequency measurements
//change parameter values from those set by CONFigure
CONF:FREQ 1.0E6, (@2)
TRIG:SOUR EXT
TRIG:SLOP POS
TRIG:COUN 2
SAMP:COUN 5
SENS:FREQ:GATE:TIME 0.005
SENS:FREQ:GATE:SOUR TIME
INIT
Using CONFigure?
The following command:
CONFigure?
returns the measurement function configured by the
last
CONFigure
or
MEASure?
command sent to the instrument. Sending
CONFigure?
after
power is cycled generates a settings conflict error until
CONFigure
or
MEASure?
is sent.
For example:
CONF:FREQ 1.0E6, (@2)
CONF?
returns (including quotes):
"FREQ +1.00000000006,+1.00000000000000E-004,(@2)"
which includes the function, expected value, (calculated) resolution, and
channel. If a channel number is not specified in the
CONFigure
or
MEASure
command, no channel is included in the return string.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...