53220A/53230A User’s Guide
85
53220A/53230A Measurements
3
Period
A standard period measurement is shown in Figure 3-4.
Figure 3-4. Standard Period Measurement.
The commands used to make period measurements are:
MEASure:PERiod? [{<expected>|MINimum|MAXimum|DEFault}
[,{<resolution>|MINimum|MAXimum|DEFault}]][,<channel>]
CONFigure:PERiod [{<expected>|MINimum|MAXimum|DEFault}
[,{<resolution>|MINimum|MAXimum|DEFault}]][,<channel>]
-
expected
is the expected input signal period.
resolution
is the desired
measurement resolution in seconds. The parameter ranges are:
expected
(channels 1 and 2): 2.8 nsec to 10 sec (default = 100 nsec)
expected
(channel 3 Option 106): 160 psec to 10 nsec (default = 2 nsec)
expected
(channel 3 Option 115): 66 psec to 3.33 nsec (default = 2 psec)
+ width
- width
t
r
t
f
period
10
90