53220A/53230A User’s Guide
111
53220A/53230A Measurements
3
Sending the command:
ABORt
aborts the measurement.
Continuous Totalize Measurement
// set up continuous totalizing, query the count after
// 10 seconds, abort the measurement after 10 additional
// seconds
*RST
// reset the counter to start from a known state
CONF:TOT:CONT (@1)
INP1:LEV 1.0
// set level threshold to 1V
INP1:SLOP NEG
// totalize negative edges
INIT
// start the measurement
(
wait 10 seconds
)
SENS:TOT:DATA?
// query count after 10 seconds
(
wait 10 seconds
)
SENS:TOT:DATA?
// query count after 20 seconds
ABOR
// stop measurement
FETC?
// retrieve final count
Notes
1. Pressing
Clear Count
on the front panel will not clear or abort continuous
measurements started from the LAN, USB, or GPIB interface. Pressing the
Back
/
Local key
, however, will clear the count when the instrument transfers
from remote to local mode.
2. Continuous totalizing must be aborted (stopped) before the count can be
retrieved from the output buffer using the
FETCh?
,
DATA:LAST?
,
DATA:REMove?
, or
R?
command (Chapter 7).
3. The
ABORt
command sent from the remote interface or changing the
counter function from the front panel stops continuous totalizing.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...