53220A/53230A User’s Guide
83
53220A/53230A Measurements
3
Frequency Ratio
A frequency ratio measurement is the measurement of two signals in which
one signal generally serves as a reference (Figure 3-3). The signals can be dif-
ferent wave shapes and applied to any 2-channel combination on the counter.
Figure 3-3. Ratio of Two Input Signal Measurements.
The commands used to make frequency ratio measurements are:
MEASure:FREQuency:RATio? [{<expected>|MINimum|MAXimum|
DEFault} [,{<resolution>|MINimum|MAXimum|DEFault}]]
[,<channel_pair>]
CONFigure:FREQuency:RATio [{<expected>|MINimum|MAXi-
mum|DEFault} [,{<resolution>|MINimum|MAXimum|DEFault}]]
[,<channel_pair>]
-
expected
is the expected
ratio
of the two input signals.
resolution
is
the desired resolution of the ratio measurement. The parameter ranges are:
expected
(Ch1/Ch2, Ch2/Ch1): 2.8E-10 to 3.5E+9
expected
(Ch1/Ch3, Ch2/Ch3 - Option 106): 1.6E-11 to 3.5
expected
(Ch1/Ch3, Ch2/Ch3 - Option 115): 6.6E-12 to 1.2
numerator channel
f
1
denominator channel
f
2
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...