53220A/53230A Measurements
Continuous
Continuous totalizing measurements count events (edges) that occur on the
input channel.
Similar to gated (timed) totalizing measurements, continuous measurements
begin when the measurement gate is opened (Figure 3-12). Rather than totaliz-
ing for a specified gate time, continuous totalizing continues indefinitely until
the measurement is aborted, or cleared and restarted from the front panel.
The command used to make continuous totalizing measurements is:
CONFigure:TOTalize:CONTinuous [<channel>]
-
channel
is counter channel 1 or 2, specified as
(@1)
or
(@2)
. (There is no
equivalent
MEASure?
command for continuous totalizing measurements.)
The command:
[SENSe:]TOTalize:DATA?
can be used during continuous totalizing or with long gate times to read the
current count.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...