53220A/53230A User’s Guide
137
53220A/53230A Input Signal Conditioning
4
The detector threshold and, therefore, the ‘on’ portion of the pulse width must
be considered when making burst measurements such as positive (on) width
and negative (off) width. This threshold is set with the command:
INPut3:BURSt:LEVel {<dB>|MINimum|MAXimum|DEFault}
INPut3:BURSt:LEVel? [{MINimum|MAXimum|DEFault}]
(query form)
The
dB
levels are -6 and -12, and are nominal (i.e. non-specified) values.
CONFigure
and
MEASure
do not change the detector threshold setting. Fol-
lowing a reset (
*RST
) or front panel preset (
Preset
), the threshold is set to
-6 dB
.
This command and burst measurement capability are only available on the
53230A
counter with
channel 3 Option 106 or 115
(6 GHz or 15 GHz micro-
wave input) and Pulse Microwave Measurement
Option 150
.
Detector Threshold Example
//configure a positive (on) width measurement on channel 3
CONFigure:PWIDth:BURSt (@3)
INP3:BURS:LEV -12
// set -12 dB detector threshold
Noise Rejection (Hysteresis)
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...