53220A/53230A User’s Guide
209
53220A/53230A Math, Graphing, and Data Logging
6
Enabling Limit Checking
Limit checking is enabled with the command:
CALCulate[1]:LIMit[:STATe] {OFF|ON}
CALCulate[1]:LIMit[:STATe]?
(query form)
ON
- enables limit checking.
OFF
disables (bypasses) limit checking. Enabling limit checking enables
both
lower and upper limit boundaries.
Setting Lower and Upper Limits
Lower and upper limits are set using the commands:
CALCulate[1]:LIMit:LOWer[:DATA] {<value>}|MINimum|MAXimum|
DEFault}
CALCulate[1]:LIMit:LOWer[:DATA]? |MINimum|MAXimum| DEFault}
(query form)
CALCulate[1]:LIMit:UPPer[:DATA] {<value>}|MINimum|MAXimum|
DEFault}
CALCulate[1]:LIMit:UPPer[:DATA]? |MINimum|MAXimum| DEFault}
(query form)
The lower limit and upper limit
value
ranges are:
-1.0E+15
to
-1.0E-15
,
0.0
,
1.0E-15
to
1.0E+15
The default limits and the limits set following a reset (
*RST
) or instrument
preset (
SYSTem:PRESet
) are 0.0.
When using limits, both a lower limit
and
an upper limit must be specified.
Any
of the following steps will prevent a “settings conflict” error (e.g. lower
limit greater than upper limit) due to the order in which limits are set:
•
setting the upper limit first
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...