90
53220A/53230A User’s Guide
3
53220A/53230A Measurements
Time Interval Measurements
The 53220A/53230A time interval measurements covered in this section
include one and two channel time interval, rise/fall time, pulse width, duty
cycle, phase, and single period.
Time Interval
N O T E
The SCPI commands listed in these examples are provided as an introduction to
how time interval measurements are made. Commands may be included even
though they specify default values - but which should be considered when modify-
ing the examples for actual use. Refer to the ‘Programmer’s Reference’ on the
Agilent 53210A/53220A/ 53230A Product Reference CD for additonal informa-
tion.
For further information on time interval measurements, see Application Note
200-3 “Fundamentals of Time Interval Measurements” also included on the CD.