53220A/53230A User’s Guide
91
53220A/53230A Measurements
3
A time interval measurement is the difference between two events, or edges,
on different waveforms or on the same waveform. A typical two-channel time
interval measurement is shown in Figure 3-6.
Figure 3-6. 2-Channel Time Interval Measurement.
The commands used to make time interval measurements are:
MEASure:TINTerval? [<channel_pair>]
CONFigure:TINTerval [<channel(s)>]
-
channel_pair
are channels in the form:
(@1),(@2)|(@2),(@1)
-
channel(s)
can be a channel pair or single channel in the form:
(@1),(@2) | (@2),(@1) | (@1)|(@2)
The input channel(s) are (default) configured for auto-leveling at 50% with a
positive slope.
When configuring time interval measurements consider the start or stop input
thresholds, input slope, or gate stop hold off settings:
Ch. 2
Ch. 1
time interval
two channel
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...