53220A/53230A User’s Guide
75
53220A/53230A Measurements
3
When shipped from the factory the measurement time out is set to 1 second.
Specifying a time of
9.9E+37
or sending the
SYSTem:SECurity:IMMediate
command disables the time out. When the time out is
disabled
, the instrument
will wait
indefinitely
for the measurement to complete.
SCPI Syntax Conventions
Programming the counters through their LAN, USB, and GPIB interface is
achieved using the Standard Commands for Programmable Instruments
(SCPI) control language. A typical example of the SCPI language syntax is
shown for the following MEASure command:
MEASure:FREQuency? [{<expected>|MINimum|MAXimum|DEFault}
[,{<resolution>|MINimum|MAXimum|DEFault}]][,<channel>]]
The programming conventions are as follows.
Command Keywords and Abbreviated Commands
Command keywords (e.g.
MEASure
,
FREQuency?
) are separated by a colon (:).
Upper case letters indicate the abbreviated spelling of the keyword or parame-
ter. Either the abbreviated form or the entire command/parameter must be
used.
Optional Keywords and Parameters
Optional (implied) keywords and parameters are indicated by square brackets
(e.g.
[,<channel>]
) and are not included when the command is sent. If an
optional parameter is not specified, a default value is used.
Separating Commands and Parameters and Linking SCPI Commands
A space must separate the last keyword and the first parameter. Commas (,)
separate multiple parameters. The ‘@’ symbol must precede
each
counter
channel number and parentheses must enclose each channel:
MEAS:FREQ:RAT? 5E6,(@2),(@1)
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...