53220A/53230A User’s Guide
163
Triggering and Gating
5
[SENSe:]FREQuency:GATE:POLarity {POSitive|NEGative}
[SENSe:]FREQuency:GATE:POLarity?
(query form)
POSitive
starts the measurement on a positive edge on the Gate In/Out BNC
or channel 1/channel 2 input, and stops the measurement on the
next
negative
edge.
NEGative
starts the measurement on a negative edge on the Gate
In/Out BNC or channel 1/channel 2 input, and stops the measurement on the
next
positive edge.
CONFigure
and
MEASure
do not change the polarity setting. Following a reset
(
*RST
) or instrument preset (
SYSTem:PRESet
or
Preset
key), the negative
edge is selected.
To set the polarity of the external gate signal:
// select the gate polarity and source
CONF:PER
SENS:FREQ:GATE:POL POS
// set polarity
SENS:FREQ:GATE:SOUR EXT
// set source
External Gate Signal Threshold
For external sources
INPut[1]
and
INPut2
(
Chan 1
and
Chan 2
softkeys), a
fixed
input threshold voltage must be specified in addition to the gate signal
polarity. This is done using the command:
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...