14
53220A/53230A User’s Guide
1
Preparation for Use
Standby power (when enabled) is provided by the line voltage or Battery
Option 300 and is used to maintain the temperature of the oven-controlled
crystal oscillator (OCXO) - Option 010. See “Applying Power”
in this chapter
for more information.
H
.
USB ‘Host’ Port
- available for transferring measurement data and
instrument configurations between the counter and a USB storage device. The
front panel port is for information transfer only. The rear panel USB port is
used for instrument (I/O) control. Data flow is covered in
Chapter 7
.
Rear Panel
The 53230A rear panel shown in Figure 1-1 includes product Option 201
(parallel rear panel inputs) and GPIB interface. The rear panel
terminals are briefly described below.
A
.
Parallel Rear Panel Inputs
- product Options 201 and 202 add parallel
inputs on the rear panel. Note that these ARE NOT separate inputs. Signals on
the center conductor of either panel’s channel BNCs are ALSO present on the
opposite panel’s center conductor. Signal path configuration is covered in
Chapter 4
.
B
.
Ext Ref In
- is the connector for providing an external reference oscillator
signal. Valid external refererence oscillator (time base) frequencies are 1, 5,
and 10 MHz.
Int Ref Out
- is the connector for accessing the counter’s internal 10 MHz
reference oscillator. The oscillator signal is a 0.5 Vrms (into 50
Ω
) sine wave.
Reference oscillator usage and configuration are covered in
Chapter 3
.
C
.
Gate In/Out
- is an
input
for external gate signals, and an
output
for
routing the counter’s internal gate to other devices. Additional information on
this connector is covered in the section “Enabling Gate Signals on the ‘Gate
in/Out’ BNC” in
Chapter 5
.
Trig In
- is the connector for supplying an external trigger signal to the
counter. Triggering is covered in
Chapter 5
.
D
.
USB and LAN
- are the standard input/output (I/O) ports. Configuration of
these ports and the GPIB interface is covered in
Chapter 2
.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...