53220A/53230A User’s Guide
161
Triggering and Gating
5
With the counter display registering:
4.999 984 583 3MHz
(11 digits - resolution enhanced)
Querying the gate time after sending this command returns:
SENS:FREQ:GATE:TIME?
+1.00000000000000E-001
(100 ms)
From Table 5-2 and assuming a 53230A counter and measurement mode
Auto
,
this corresponds to 11 digits of resolution. To achieve a desired resolution
(digits) in terms of gate time, locate the number of digits in Table 5-2 and
select the corresponding gate time:
SENS:FREQ:GATE:TIME 100e-3
// set gate time = 100 ms
SENS:FREQ:GATE:SOUR TIME
// set gate source
As another example, assume a measurement requiring 6-digit resolution of a
5 ns (200 MHz) signal. Recalling that
Log
10
(
expected value
) -
Log
10
(
resolu-
tion
) is approximately equal to digits of resolution, the measurement can be
configured as:
MEAS:PER? 5E-9, 5E-15, (@1) // per meas of 200 MHz signal
A typical measurement using this configuration might return:
+5.00010899135045E-009
With the counter display registering:
5.000 10nsec
(6 digits)
Querying the gate time after sending this command returns:
SENS:FREQ:GATE:TIME?
+1.00000000000000E-005
(10 us - no resolution enhancement)
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...