86
53220A/53230A User’s Guide
3
53220A/53230A Measurements
resolution
(all channels): 1.0E-15 *
expected
to 1.0E-5*
expected
(default
resolution
corresponds to a 0.1s gate time)
-
channel
is counter channel 1, 2, or 3 specified as
(@1)
,
(@2)
, or
(@3)
.
Period Measurement Examples
// using MEASure? - measure the period of a 100 ns (10 MHz)
// signal with 12-digit resolution
*RST
// reset to start from known state
MEAS:PER? 100E-9, 1E-12, (@1)
----------------------------------------------------------
// using CONFigure - measure the period of a 100 ns (10 MHz)
// signal with 12-digit resolution
*RST
// reset to start from known state
SYST:TIM 1.0
// 1s measurement timeout (per meas)
CONF:FREQ 100E-9, 1E-12, (@1)
SAMP:COUN 10 // take 10 readings
READ?
Notes
1. See Chapter 5 for information on triggering and number of readings per trig-
ger (sample count).
2. See Chapter 4 for additional information on counter threshold levels and on
configuring the input signal path.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...