96
53220A/53230A User’s Guide
3
53220A/53230A Measurements
// using CONFigure - measure
rise time
on the wave segment
// from -1.75V to +750 mV
*RST
// reset to start from known state
CONF:RTIM -1.75 V, 750 MV, (@1)
INP:COUP DC // set AC coupling
INP:IMP 50
// set input impedance to 50 ohm
READ?
----------------------------------------------------------
// using MEASure? - measure
fall time
using 15% and 65%
// references
*RST
// reset to start from known state
MEAS:FTIM? 15PCT, 65PCT, (@2)
----------------------------------------------------------
// using CONFigure - measure
fall time
on the wave segment
// from +1.25V to -1.4V
*RST
// reset to start from known state
CONF:FTIM -1.4V, 1.25V, (@1)
INP:COUP DC // set DC coupling
INP:IMP 50
// set input impedance to 50 ohm
READ?
Notes
1. Lower and upper references can be specified as percentages or as absolute
levels in either the
CONFigure
or
MEASure?
command. For either rise time or
fall time measurements, the lower reference is specified first.
2. Absolute references are relative to the amplitude of the signal and must take
into consideration the counter’s input coupling and impedance settings.
3. See Chapter 4 for information on configuring the input signal path which
includes coupling and impedance.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...