53220A/53230A User’s Guide
3
53220A/53230A Measurements
Figure 3-12. Gated and Continuous Totalizing.
The commands used to make gated (timed) totalizing measurements are:
MEASure:TOTalize:TIMed? [{<gate_time>|MINimum|MAXimum|
DEFault}] [,<channel>]
CONFigure:TOTalize:TIMed [{<gate_time>|MINimum|MAXimum|
DEFault}] [,<channel>]
-
gate_time
specifies the time input events on the counter channel are total-
ized. The range for
gate_time
is:
53220A: 100
µ
s to 1000s (10
µ
s resolution) or +9.9E+37 (INFinity)
53230A: 1
µ
to 1000s (1
µ
s resolution) or +9.9E+37 (INFinity)
The default
gate_time
for both instruments is 0.100s.
-
channel
is counter channel 1 or 2, specified as
(@1)
or
(@2)
.
Gate open
Gate closed
gate
input
gate time
...
events
(gated)
(continuous)