53220A/53230A User’s Guide
69
53220A/53230A Measurements
3
The presence of a valid external reference can be determined programmati-
cally with the command:
[SENSe:]ROSCillator:EXTernal:CHECk ONCE
Prior to sending the command,
SENSe:ROSCillator:SOURce EXTernal
and
SENSe:ROSCillator:SOURce:AUTO OFF
must be set.
The following example checks for an external reference signal applied to the
Ext Ref In
connector.
SENS:ROSC:SOUR EXT
// set source, also disables
// automatic source selection
SENS:ROSC:EXT:CHEC ONCE
// check for valid signal
SYST:ERR?
// read the error queue
The error queue can be read from the front panel as follows:
The status of the external reference (time base) is also monitored by the
counter’s Questionable Data register. The register can be queried with the
command:
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...