166
53220A/53230A User’s Guide
5
Triggering and Gating
INFinity|DEFault}
[SENSe:]TOTalize:GATE:TIME? {MINimum|MAXimum|DEFault}
(query form)
CONFigure
and
MEASure
automatically set the gate source to
TIME,
and the
gate time to the value of their
gate_time
parameter.
To specify a gate time and source directly for totalizing:
// set gate time and source
SENS:TOT:GATE:TIME 10
// set gate time = 10s
SENS:TOT:GATE:SOUR TIME
// set gate source
External Gate Sources
Gate sources
EXTernal, INPut[1],
and
INPut2
are external sources.
EXTernal
is the counter’s rear panel ‘Gate In/Out’ BNC, and
INPut[1]/INPut2
(represented by softkeys
Chan 1
and
Chan 2
) are the
counter’s channel 1 and channel 2 inputs.
External Gate Signal Polarity
When using the external gate sources, the polarity of the gate signal and there-
N O T E
OUTput:STATe
OFF
must be set when using gate source
EXTernal
. See
“Enabling Gate Signals on the ‘Gate In/Out BNC” for more information and for
information on using gate signals to synchronize other instruments.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...