198
53220A/53230A User’s Guide
6
53220A/53230A Math, Graphing, and Data Logging
Scaling function results must be in the range of
-1.0E+24 to -1.0E-24
,
0.0
, or
+1.0E-24 to 1.0E+24
. Results outside these limits will be
replaced with
-9.9E+37
(negative infinity),
0
, or
+9.9E+37
(positive infin-
ity).
Following a reset (
*RST
) or instrument preset (
SYSTem:PRESet
), the scaling
function is set to
NULL
.
Scale Reference Value
The
NULL
,
PCT
,
PPM
, and
PPB
scaling functions require a reference value. For
PCT
,
PPM
, and
PPB
the value cannot be ‘0’
.
The reference can be obtained auto-
matically or specified directly with the commands:
CALCulate[1]:SCALe:REFerence:AUTO {OFF|ON}
CALCulate[1]:SCALe:REFerence:AUTO?
(query form)
CALCulate[1]:SCALe:REFerence {<reference>|MINimum|MAXimum|
DEFault}
CALCulate[1]:SCALe:REFerence? [{MINimum|MAXimum|DEFault}]
(query form)
ON
-
automatically selects the first measurement as the reference for all subse-
quent readings in the reading count (trigger count x sample count).
OFF
dis-
ables automatic selection and requires the reference to be specified directly.
Following a reset (
*RST
) or instrument preset (
SYSTem:PRESet
), automatic
reference selection is enabled (
ON
).
<
reference
> -
specifies the reference value directly. The reference is used
for all readings in the reading count.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...