124
53220A/53230A User’s Guide
4
53220A/53230A Input Signal Conditioning
Input Range
The signal operating ranges (input ranges) of the 53220A/53230A counter are
±
5.0V, ± 50V,and ± 500V and are dependent on the probe factor. The range is
set using the command:
INPut[{1|2}]:RANGe {<range>|MINimum|MAXimum|DEFault}
INPut[{1|2}]:RANGe? [{MINimum|MAXimum|DEFault}]
(query form)
When a 1:1 probe factor is selected (see “Setting the Probe Factor), the avail-
able ranges are 5.0V and 50.0V. When a 10:1 probe factor is selected the avail-
able ranges are 50.0V and 500.0V.
CONFigure
and
MEASure
do not change the input range setting. Following a
reset (
*RST
) or a front panel preset (
Preset
), the range is set to 5.0V.
Using Auto Scale
For input signals greater than 100 Hz,
Auto Scale
detects the signal on the input
channel and sets the range to 5V or 50V depending on the amplitude.
Auto Scale
is
not
the same as an instrument preset which sets a default range of 5V.
Range Example
//period measurement of expected 10 MHz signal,
//maximum resolution, use channel 1
CONF:PER 0.1E-6,MAX,(@1)
INP:IMP 1E6
//set impedance to 1 Mohm
INP:RANG 50
//set 50V range
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...