280
53220A/53230A User’s Guide
A
53220A/53230A Error Messages
-257
File name error;
file too large
The file uploaded to the counter is greater than
1 GByte.
-257
File name error;
folder not empty
When deleting a folder from an I/O (remote) inter-
face, the folder must be empty - no sub-folders or
files. Un-empty folders CAN be deleted from the
front panel.
-257
File name error;
unknown file extension
Depending on the memory operation and file type,
valid file extensions are .csv, .dat, .sta, and .prf.
Device Specific Errors
-310
System Error;
internal software error
At power-on, the firmware is unable to obtain
information from the instrument’s non-volatile
memory to complete initialization. Cycle power
again. Contact Agilent if the error remains.
-310
System Error;
failed to erase calibration data
in PIC EEProm
For system errors related to erasing, reading, or
writing calibration data, verify that the instrument
has the latest firmware revision installed. Firm-
ware updates can be located beginning at the
“Technical Support” tab at
www.agilent.com/find/53220A or
www.agilent.com/find/53230A. If the error condi-
tion remains, contact Agilent.
-310
System Error;
failed to erase system infor-
mation in PIC EEProm
For system errors related to erasing, reading, or
writing system information, verify that the instru-
ment has the latest firmware revision installed.
Firmware updates can be located beginning at the
“Technical Support” tab at
www.agilent.com/find/53220A or
www.agilent.com/find/53230A. If the error condi-
tion remains, contact Agilent.
-310
System Error;
I2C Comms Failure
I2C related errors if they occur, will occur most
often during boot up when power is cycled. If this
type of error occurs cycle power again. If the error
remains, contact Agilent with a complete descrip-
tion of the error message.
-313
Calibration memory lost;
memory corruption detected
The instrument was unable to reference calibra-
tion data in non-volatile memory. Perform a recali-
bration.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...