70
53220A/53230A User’s Guide
3
53220A/53230A Measurements
STATus:QUEStionable:EVENt?
A value of
+1024
(bit 10) indicates an invalid frequency (time base) error.
Reading the register
clears all bits
in the register. See Chapter 8 for informa-
tion on the counter’s status system.
Standby Power to the Reference Oscillator (Option 010)
Standby power to maintain the operating temperature of the counter’s
internal OCXO reference oscillator (Option 010) is provided by the the line
voltage or from Battery Option 300. Standby power is enabled and disabled
using the command:
[SENSe:]ROSCillator:INTernal:POWer:STANdby {OFF|ON}
[SENSe:]ROSCillator:INTernal:POWer:STANdby?
(query form)
-
On
enables standby power which maintains the temperature of the OCXO
when the front panel power button is switched ‘off’ (standby). If line power is
removed from the instrument and Battery Option 300 is installed
and
enabled
, standby power is supplied by the battery. Battery-supplied standby
power is 24 hours.
-Off
disables standby power from either the line voltage or battery to the
internal oscillator when the front panel power button is switched ‘off’.
With standby power
Off
, cycling power on the instrument also cycles power to
the oscillator.
This may affect the oscillator’s calibrated accuracy, and sta-
bility may not occur until the specified warm-up time (45 minutes) is
reached.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...