170
53220A/53230A User’s Guide
5
Triggering and Gating
Time Interval Measurements
For time interval measurements, the command used to set/change the gate
source is:
[SENSe:]TINTerval:GATE:SOURce {IMMediate|EXTernal|
ADVanced}
[SENSe:]TINTerval:GATE:SOURce?
(query form)
- gate source
IMMediate
uses an internal gate signal which starts the mea-
surement on the first event (edge/level) defined by the
INPut
subsystem, and
ends on the next (stop) event defined. The
CONFigure
commands set the time
interval gate source to
IMMediate
.
- gate source
EXTernal
is the counter’s rear panel ‘Gate In/Out’ BNC. Using
an external gate, the measurement begins on the first start event
after
the
external gate is received. The measurement ends following the stop event. The
start and stop events (edge/level) are defined by the
INPut
subsystem.
N O T E
OUTput:STATe
OFF
must be set when using gate source
EXTernal
. See
“Enabling Gate Signals on the ‘Gate In/Out BNC” for more information and for
information on using gate signals to synchronize other instruments.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...