53220A/53230A User’s Guide
185
Triggering and Gating
5
- holdoff source
TIME
holds off the gate stop (close) for a period set by the
command:
[SENSe:]GATE:STOP:HOLDoff:TIME {<time>|MINimum|MAXimum|
INFinity|DEFault}
[SENSe:]GATE:STOP:HOLDoff:TIME?
(query form)
Note for frequency, frequency ratio, average period, PRF, and PRI measure-
ments, the minimum gate stop (close) holdoff time is 100 usec for the 53220A,
or 1 usec for the 53230A.
INFinity
or
+9.9E+37
is accepted
only
for the TOTalize measurement func-
tion. When selected, the gate remains open until an ABORt or *RST command
or a device clear is received.
Gate Stop Source
The gate stop source determines when the measurement gate is closed
follow-
ing
the stop hold off.
N O T E
Setting a gate stop source does not apply to rise/fall time, duty cycle, pulse width,
phase, and single period measurements. These measurements complete auto-
matically on the next appropriate edge on the input channel.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...