53220A/53230A User’s Guide
4
53220A/53230A Input Signal Conditioning
be enabled within the signal path to eliminate noise introduced by higher-fre-
quency components of the input signal.
The bandwidth filter is switched into the signal path using the command:
INPut[{1|2}]:FILTer[:LPASs][:STATe] {OFF|ON}
INPut[{1|2}]:FILTer[:LPASs][:STATe]?
(query form)
On
- enables the filter.
OFF
- disables the filter.
Figure 4-3 shows the effects on the instrument’s
measurable frequency range
when the filter is enabled.
Figure 4-3. Measureable Frequency Range with Bandwidth Filter Enabled.
From the figure:
•
DC coupling = 1 mHz - 100 kHz
•
AC coupling = 10 Hz - 100 kHz
CONFigure
and
MEASure
do not change the filter setting. Following a reset
(
*RST
) or front panel preset (
Preset
), the low-pass filter is disabled (off).
100 kHz
10 Hz
DC
AC
1 mHz
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...