53220A/53230A User’s Guide
225
53220A/53230A Math, Graphing, and Data Logging
6
Viewing Trend Charts
At power-on or following a reset (
*RST
) or instrument preset (
Preset
), numeric
data is displayed. The display can be changed to a trend chart by changing the
display mode using the command:
DISPlay[:WINDow]:MODE {NUMeric|HISTogram|TCHart}
DISPlay[:WINDow]:MODE?
(query form)
The display mode (
TCHart
) is the
only
trend chart parameter that can be set
programmatically (i.e. using SCPI commands). All other trend chart parame-
ters are set using
the front panel
.
Pressing the
Graph
key followed by the
Trend Chart
softkey starts the trend chart
from the front panel.
Trend charts correspond to
all
readings in reading memory for a given trigger
count (
TRIGger:COUNt
) and sample count (
SAMPle:COUNt
). If triggering is
continuous (i.e. internal), the trend chart continually updates.
Trend Chart Configuration
Figure 6-9 shows the softkey menus associated with trend chart configuration
and control of the trend window.
Содержание 53220A
Страница 1: ...Agilent Technologies Agilent 53220A 53230A 350 MHz Universal Frequency Counter Timer User s Guide...
Страница 10: ...10 53220A 53230A User s Guide Contents...
Страница 34: ...34 53220A 53230A User s Guide 1 Preparation for Use...
Страница 62: ...62 53220A 53230A User s Guide 2 53220A 53230A Software Installation and Interface Configuration...
Страница 142: ...142 53220A 53230A User s Guide 4 53220A 53230A Input Signal Conditioning...
Страница 190: ...190 53220A 53230A User s Guide 5 Triggering and Gating...
Страница 268: ...268 53220A 53230A User s Guide 8 Instrument Status Figure 8 1 The 53220A 53230A Status System...
Страница 274: ...274 53220A 53230A User s Guide 8 Instrument Status...
Страница 288: ...288 53220A 53230A User s Guide A 53220A 53230A Error Messages...