background image

5314A

UNIVERSAL COUNTER

OPERATING AND SERVICE MANUAL

This  manual applies directly to HP Model 5314A

having serial number prefix  1828A and 

below.

This manual, with enclosed “Manual Changes” sheet,
applies to HP Model 5314A having serial number
prefixes as listed on the “Manual Changes” sheets.

Copyright H E W L E T T - P A C K A R D   C O M P A N Y 

1978

5301 STEVENS CREEK BLVD., SANTA CLARA,  CALIF.  95050

MANUAL PART NUMBER 05314-90001
Microfiche Part Number 05314-90002

Printed: SEPT 1978

HEWLETT

Содержание 5314A

Страница 1: ...ow This manual with enclosed Manual Changes sheet applies to HP Model 5314A having serial number prefixes as listed on the Manual Changes sheets Copyright H E W L E T T P A C K A R D C O M P A N Y 197...

Страница 2: ...gh this instrument has been designed in accordance with international safety standards this manual contains information cautions and warnings which must be followed to ensure safe operation and to ret...

Страница 3: ...h must be followed by the user to ensure safe operation and keep the HP 5314A in safe operating condition 1 8 INSTRUMENTS COVERED BY MANUAL I 9 If the serial number prefix of your HP 5314A is lower th...

Страница 4: ...in decade steps of N for N l to 1000 Resolution LSD B Trigger Error X FREQUENCY A N Accuracy 1 count of A B Trigger Error X FREQUENCY A TOTALIZE A Range 10 Hz to 10 MHz Resolution t 1 count of input...

Страница 5: ...for the HP 5314A They are a Temperature Compensated Crystal Oscillator TCXO Option 001 and a battery pack for portable operation Option 002 I 17 The seven functions are Frequency Single Shot Period P...

Страница 6: ...for use during performance tests adjustments and troubleshooting Substitute test equipment may be used if it meets the required characteristics listed in the table Table 7 3 Recommended Test Equipmen...

Страница 7: ...Time Interval Measurements 3 10 The counter measures time intervals from Channel A to Channel B that is Channel A starts the measurement and Channel B stops the measurement Time between points on a s...

Страница 8: ...is the same Description numbers match the numbers on the illustration 3 22 OPERATOR S MAINTENANCE 3 23 There is no operator s maintenance for the HP 5314A All maintenance should be per formed by qual...

Страница 9: ...2 battery charging circuit in the STBY position Front panel window allowing access to the oscillator frequency adjust ment capacitor AlC2 In HOLD switch IN the measurement except for totalize in progr...

Страница 10: ...uses the HP 5314A to make 100 measurements and display the average ratio This switch does not improve accuracy beyond 100 nanoseconds for time interval measurements 0 I 10 Hz N 10 In frequency 10 Hz t...

Страница 11: ...n with the attenuator switch 0 to select the relative voltage at which triggering occurs Approximately t350 millivolts is the amount varied The input amplifiers are ac coupled The actual dc level of t...

Страница 12: ...ifi 6 Select either 1 Hz or 10 Hz NOTE 10 Hz resolution for frequencies between 10 Hz and 10 MHz may also be used For frequencies higher than 10 MHz the 100MHz 1O Hz switch must be pressed IN NOTE The...

Страница 13: ...acy and amplitude 1 Set line switch to the ON position 2 Set COM A SEP switch to SEP position 3 Connect the input signal to INPUT A jack 4 Press PER A T l A B switch IN Make sure the blue shift key is...

Страница 14: ...position 3 Press PER A T I A B switch 0 bottom function of switch 0 IN Press the blue shift key IN position This selects the 4 Connect the start time interval signal to the INPUT A jack Connect the st...

Страница 15: ...EP position 3 Place the two function switches and in the OUT position This places the HP 5314A in the Ratio A to B function 4 Connect the higher frequency signal to the INPUT A jack D Connect the lowe...

Страница 16: ...accuracy and amplitude 1 Set line switch to the ON position 2 Set COM A SEP switch 0 to SEP position 3 Press the FREQ A START A switch IN 4 Press the blue shift key IN position This selects the bottom...

Страница 17: ...n the IN position The HP 5314A should display with the overflow LED 0 ON and the instrument gating once every second 4 Place resolution switch 10 Hz N 10 in the IN position The HP 5314A should display...

Отзывы: