Helwett Packard 5314A Скачать руководство пользователя страница 10

Model 5314A

Operation

NOTE

0

SHIFT KEY

 1  Hz/N=100

There are two additional functions which are selected using
combinations of switches @and@. These two functions
are Self-Check and Ratio A to B. For self-check mode, place

both function switches

 and @in the IN position. The

instrument is now making a frequency measurement on the
internal 10 MHz time base. Activating switches
causes 10 MHz to be displayed. Activating
causes 100 MHz to be displayed. Resolution selection
switches can now be checked for

Ratio A to B, place both function switches
OUT position. For more details on
paragraph 3-12 and 

Figure 

3-7.

IN/OUT position determines the function selected by keys 

 IN

position selects the bottom row functions. OUT position selects the upper
row functions.

NOTE

The following three switches

are dual purpose.

Depending on the function selected (Frequency, Period,
Ratio, etc.) the switches either represent the resolution and
bandwidth (gate time) or the sample size (N samples).

In frequency 

(10 

Hz to 

10 MHz), 

this switch, 

when 

IN, 

gives a display with a

1 Hz resolution (1 second gate time). For frequencies between 10 MHz and
100 MHz, see the explanation for switch@. In period, this switch, when

IN, 

causes the HP 5314A to measure 100 periods and display the average

value 

in microseconds. 

In ratio, this switch when 

IN, 

causes the 

HP 5314A 

to

make 100 measurements and display the average ratio. This switch does 

not

improve accuracy beyond 100 nanoseconds 

for time interval

measurements!

10  Hz/N=10

In frequency (10 Hz to 10 MHz), this switch, when 

IN, 

gives a display with a

10 Hz resolution (100 millisecond gate time). For frequencies between
10 MHz and 100 MHz, see the explanation 

for switch@ 

 In period, this

switch, when IN, causes the 

HP 5314A 

to 

measure 10 periods and display

the average value 

in microseconds.

In ratio, this switch, when 

IN, 

causes the HP 5314A to make 10 measure-

ments and display the average ratio. This switch does 

not 

improve accuracy

beyond 100 nanoseconds 

for time interval measurements!

NOTE

There is another resolution available using 

 in addition to the two resolutions called out on the

front panel. It is 0.1 Hz N = 1 0 0 0  This is generated when
switches

 

are in the OUT position. In fre-

quency (10 z to 10 MHz), with these three switches out,
the 

HP 5314A 

gives a display with 0.1 Hz resolution (10

second 

gate time). For frequencies between 

 MHz and

100 MHz, see the explanation 

for switch 

 In period,

with these three switches out, the HP 5314A measures 1000

periods and displays the average value in microseconds. In
ratio, with these three switches out, the HP 5314A makes
1000 measurements and displays the average ratio. This

switch combination does 

not 

improve accuracy beyond

100 

nanoseconds 

for time interval measurements.

Figure 

3-2. Front 

Panel Controls and Connectors (Continued)

3-5

Содержание 5314A

Страница 1: ...ow This manual with enclosed Manual Changes sheet applies to HP Model 5314A having serial number prefixes as listed on the Manual Changes sheets Copyright H E W L E T T P A C K A R D C O M P A N Y 197...

Страница 2: ...gh this instrument has been designed in accordance with international safety standards this manual contains information cautions and warnings which must be followed to ensure safe operation and to ret...

Страница 3: ...h must be followed by the user to ensure safe operation and keep the HP 5314A in safe operating condition 1 8 INSTRUMENTS COVERED BY MANUAL I 9 If the serial number prefix of your HP 5314A is lower th...

Страница 4: ...in decade steps of N for N l to 1000 Resolution LSD B Trigger Error X FREQUENCY A N Accuracy 1 count of A B Trigger Error X FREQUENCY A TOTALIZE A Range 10 Hz to 10 MHz Resolution t 1 count of input...

Страница 5: ...for the HP 5314A They are a Temperature Compensated Crystal Oscillator TCXO Option 001 and a battery pack for portable operation Option 002 I 17 The seven functions are Frequency Single Shot Period P...

Страница 6: ...for use during performance tests adjustments and troubleshooting Substitute test equipment may be used if it meets the required characteristics listed in the table Table 7 3 Recommended Test Equipmen...

Страница 7: ...Time Interval Measurements 3 10 The counter measures time intervals from Channel A to Channel B that is Channel A starts the measurement and Channel B stops the measurement Time between points on a s...

Страница 8: ...is the same Description numbers match the numbers on the illustration 3 22 OPERATOR S MAINTENANCE 3 23 There is no operator s maintenance for the HP 5314A All maintenance should be per formed by qual...

Страница 9: ...2 battery charging circuit in the STBY position Front panel window allowing access to the oscillator frequency adjust ment capacitor AlC2 In HOLD switch IN the measurement except for totalize in progr...

Страница 10: ...uses the HP 5314A to make 100 measurements and display the average ratio This switch does not improve accuracy beyond 100 nanoseconds for time interval measurements 0 I 10 Hz N 10 In frequency 10 Hz t...

Страница 11: ...n with the attenuator switch 0 to select the relative voltage at which triggering occurs Approximately t350 millivolts is the amount varied The input amplifiers are ac coupled The actual dc level of t...

Страница 12: ...ifi 6 Select either 1 Hz or 10 Hz NOTE 10 Hz resolution for frequencies between 10 Hz and 10 MHz may also be used For frequencies higher than 10 MHz the 100MHz 1O Hz switch must be pressed IN NOTE The...

Страница 13: ...acy and amplitude 1 Set line switch to the ON position 2 Set COM A SEP switch to SEP position 3 Connect the input signal to INPUT A jack 4 Press PER A T l A B switch IN Make sure the blue shift key is...

Страница 14: ...position 3 Press PER A T I A B switch 0 bottom function of switch 0 IN Press the blue shift key IN position This selects the 4 Connect the start time interval signal to the INPUT A jack Connect the st...

Страница 15: ...EP position 3 Place the two function switches and in the OUT position This places the HP 5314A in the Ratio A to B function 4 Connect the higher frequency signal to the INPUT A jack D Connect the lowe...

Страница 16: ...accuracy and amplitude 1 Set line switch to the ON position 2 Set COM A SEP switch 0 to SEP position 3 Press the FREQ A START A switch IN 4 Press the blue shift key IN position This selects the bottom...

Страница 17: ...n the IN position The HP 5314A should display with the overflow LED 0 ON and the instrument gating once every second 4 Place resolution switch 10 Hz N 10 in the IN position The HP 5314A should display...

Отзывы: