Model 5314A
General 

Information

Table  1-1. Specifications

INPUT CHARACTERISTICS

Range:

Channel A 10 Hz 

to 100 

MHz

Channel B 10 Hz to 2.5 MHz

Sensitivity:

Channel A:

25 mV rms to 100 MHz
75 mV peak-to-peak minimum pulse with 5 ns

Channel B:

25 mV rms to 2.5 MHz
75 mV peak-to-peak minimum pulse width of 50 ns

Coupling: AC
Impedance: 

 NOMINAL 

shunted by less than

30 pF

Attenuator: 

Xl or 

X20 NOMINAL 

(A Channel only)

Trigger Level:

Continuously variable 

 mV times attenuator

setting around average value 

OS 

signal.

Slope: Independent selection of + or - slope
Channel Input: Selectable SEPARATE or COMMON A
Damage Level:

Xl : 

DC to 100 kHz

350V (DC + peak AC)

100 kHz to 5 MHz 2.5 X 

 X Hz Product

Above 5 MHz 5V rms

x20: DC to 

1 MHz 350V (DC + Peak AC)

1 MHz to 50 MHz 2.5 X 

 X Hz 

Product

Above 50 MHz 5V rms

FREQUENCY (A)

Range:

10 

Hz to 10 MHz direct count

10 Hz to 100 MHz prescaled by 10

LSD Displayed: Direct count 0.1 Hz, 1 Hz, 10 Hz switch

selectable. Prescaled 10 Hz, 100 Hz, 1 kHz switch
selectable.

Resolution:   LSD

Accuracy:   LSD   (time base error) X

 FREQ

PERIOD (A)

Range: 10 Hz to 2.5 MHz
LSD Displayed:

100 

ns for 

N=l

 to 

1000 

in decade steps of 

N

N

Resolution:

 LSD   1.4 X

Trigger Error

N

Accuracy

 LSD   1.4  X Trigger Error

N

 (time base error) 

PER

TIME INTERVAL (A TO B)

Range: 250 ns to 1 s
LSD Displayed: 

100 

ns

Resolution: 

 LSD   START 

Trigger Error   STOP

Trigger Error

Accuracy: 

 LSD   START 

Trigger Error   STOP

Trigger Error   (time base error) X T1

Time Interval measurements require an arming signal 

for

both the 

START 

and STOP Channels.

(See Paragraph 3-11 .)

RATIO

Range:

10 

Hz to 10 MHz 

Channel 

A

10 Hz to 2.5 MHz 

Channel B

LSD Displayed:

1 part in

 X N in decade steps 

of 

for 

N=l 

to 1000

Resolution:

 LSD   (B Trigger Error X FREQUENCY A)/N

Accuracy:

 1 

count of A   B Trigger Error 

X FREQUENCY A

TOTALIZE  (A)

Range: 10 Hz to 10 MHz
Resolution: 

 1 

count of input

GENERAL

Check: Counts internal 10 MHz Oscillator
Display: 7-digit amber LED display with gate and over-

flow indication.

Maximum Ssmple Rate: 5 readings per second.
Operating Temperature: 

 to 50°C

Power Requirement:

100/120/230/240V  rms 

+5%, -IO%, 48-66 Hz; 10 VA

maximum.

Weight: 2.0 kg (4.4 

Ibs.)

Dimension: 238 mm 

wide X

 98 

mm high 

X276 

mm long

 X 

 X 

 in.)

TIME BASE

Frequency: 

10 MHz

Aging Rate: <3 parts in 

 per month

Temperature: 

 part in 

 0 °  to 50°C

line Voltage: 

 part in 

 for 

 variation.

OPTIONS

Option 001: High Stability Time Base (TCXO)

Frequency: 10 MHz

Aging Rate: 

 part in 

 per month

Temperature: 

 part in 

 0° to 40°C

line Voltage: 

 part in 

 for 

 variation

Option 002: Battery

Type: Rechargeable lead-acid (sealed)

Capacity: TYPICALLY 8 hour of continuous operation

at 

25°C.

Recharging Time: TYPICALLY 16 hours to 98% of full

charge, instrument nonoperating. Charging cir-
cuitry included with option. Batteries not charged
during instrument operation.

Battery Voltage Sensor: Automatically shuts instru-

ment off when low battery condition exists.

line Failure Protection: 

Instrument

 automatically

switches to batteries in case of line failure.

Weight: Option 002 adds I.5 kg (3.3 Ibs.) to weight

of instrument.

DEFINITIONS

Resolution: Smallest discernible change of measure-

ment result due to a minimum change in the input.

Accuracy: Deviation from the actual value as fixed by

universally accepted standard of frequency and

time.

Trigger Error:

 

 + 

Input Slew Rate at Trigger Point 

(rms)

i

Where 

 is the rms noise of the input 

for a 100

MHz bandwidth on Channel A and a 10 MHz
bandwidth on Channel 

B

LSD: Least Significant Digit.

Содержание 5314A

Страница 1: ...ow This manual with enclosed Manual Changes sheet applies to HP Model 5314A having serial number prefixes as listed on the Manual Changes sheets Copyright H E W L E T T P A C K A R D C O M P A N Y 197...

Страница 2: ...gh this instrument has been designed in accordance with international safety standards this manual contains information cautions and warnings which must be followed to ensure safe operation and to ret...

Страница 3: ...h must be followed by the user to ensure safe operation and keep the HP 5314A in safe operating condition 1 8 INSTRUMENTS COVERED BY MANUAL I 9 If the serial number prefix of your HP 5314A is lower th...

Страница 4: ...in decade steps of N for N l to 1000 Resolution LSD B Trigger Error X FREQUENCY A N Accuracy 1 count of A B Trigger Error X FREQUENCY A TOTALIZE A Range 10 Hz to 10 MHz Resolution t 1 count of input...

Страница 5: ...for the HP 5314A They are a Temperature Compensated Crystal Oscillator TCXO Option 001 and a battery pack for portable operation Option 002 I 17 The seven functions are Frequency Single Shot Period P...

Страница 6: ...for use during performance tests adjustments and troubleshooting Substitute test equipment may be used if it meets the required characteristics listed in the table Table 7 3 Recommended Test Equipmen...

Страница 7: ...Time Interval Measurements 3 10 The counter measures time intervals from Channel A to Channel B that is Channel A starts the measurement and Channel B stops the measurement Time between points on a s...

Страница 8: ...is the same Description numbers match the numbers on the illustration 3 22 OPERATOR S MAINTENANCE 3 23 There is no operator s maintenance for the HP 5314A All maintenance should be per formed by qual...

Страница 9: ...2 battery charging circuit in the STBY position Front panel window allowing access to the oscillator frequency adjust ment capacitor AlC2 In HOLD switch IN the measurement except for totalize in progr...

Страница 10: ...uses the HP 5314A to make 100 measurements and display the average ratio This switch does not improve accuracy beyond 100 nanoseconds for time interval measurements 0 I 10 Hz N 10 In frequency 10 Hz t...

Страница 11: ...n with the attenuator switch 0 to select the relative voltage at which triggering occurs Approximately t350 millivolts is the amount varied The input amplifiers are ac coupled The actual dc level of t...

Страница 12: ...ifi 6 Select either 1 Hz or 10 Hz NOTE 10 Hz resolution for frequencies between 10 Hz and 10 MHz may also be used For frequencies higher than 10 MHz the 100MHz 1O Hz switch must be pressed IN NOTE The...

Страница 13: ...acy and amplitude 1 Set line switch to the ON position 2 Set COM A SEP switch to SEP position 3 Connect the input signal to INPUT A jack 4 Press PER A T l A B switch IN Make sure the blue shift key is...

Страница 14: ...position 3 Press PER A T I A B switch 0 bottom function of switch 0 IN Press the blue shift key IN position This selects the 4 Connect the start time interval signal to the INPUT A jack Connect the st...

Страница 15: ...EP position 3 Place the two function switches and in the OUT position This places the HP 5314A in the Ratio A to B function 4 Connect the higher frequency signal to the INPUT A jack D Connect the lowe...

Страница 16: ...accuracy and amplitude 1 Set line switch to the ON position 2 Set COM A SEP switch 0 to SEP position 3 Press the FREQ A START A switch IN 4 Press the blue shift key IN position This selects the bottom...

Страница 17: ...n the IN position The HP 5314A should display with the overflow LED 0 ON and the instrument gating once every second 4 Place resolution switch 10 Hz N 10 in the IN position The HP 5314A should display...