182
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
8
Replace the FLF board and fix the screws (A and B) and
connectors (C and D).
9
Check that the safety switch is positioned correctly. Look
through board and casting.
10
Reconnect the wires to the connector. Use a flat screw
driver to release the wire socket.
11
Put the wires into the released contacts using the
tweezers.
Order of wires (left to right):
Next Steps:
•
Reinstall the optical unit as described in
•
If you replaced a G1321-66501 FLF board by a G1321-66511 FLF board, update the FLD firmware to revision A.03.7x or
above (to make it compatible).
A
B
C
D
green
empty
brown
yellow
white
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......