First Steps with the Fluorescence Detector
2
1100 Series FD Reference Manual
55
5
Using the settings and the timetable (from previous page), do a second run
for the evaluation of the optimal excitation wavelength. See
.
Table 13
Peak #
Time
Emission Wavelength
1
5.3 min
330 nm
2
7.2 min
330 nm
3
7.6 min
310 nm
4
8.6 min
360 nm
5
10.6 min
445 nm
6
11.23 min
385 nm
Figure 21
Detector Settings for Excitation Scan
DO NOT select
additional excitation
wavelengths (B, C,
D). Doing so will
increase the scan
time and will lower
the performance