92
1100 Series FD Reference Manual
4
Troubleshooting and Test Functions
FLF Board not found
The FLF board could not be found by the main board (FLM). This message
comes together with some other message generated on the FLF board (e.g.
Leak, ...).
Probable Causes
•
FLF board not connected to the FLM board.
•
defective FLF board.
•
defective FLM board.
Suggested Actions
✔
Check the connection between the FLF board and the FLM board.
✔
Exchange the FLF board.
✔
Exchange the FLM board.