
Installing the Fluorescence Detector
1
1100 Series FD Reference Manual
23
8
Connect the APG remote cable (optional) for non-Agilent 1100 Series
instruments.
9
Turn on power by pushing the button at the lower left hand side of the
detector. The status LED should be green.
Figure 6
Rear View of Detector
Analog signal
APG remote
RS-232C
CAN
GPIB
Power
Configuration switch
Interface board
LAN or BCD/EXT
Security lever
N O T E
The detector is turned on when the line power switch is pressed and the green indicator
lamp is illuminated. The detector is turned off when the line power switch is protruding and
the green light is off.
WA R N I N G
To disconnect the detector from line, unplug the power cord. The power supply still
uses some power, even if the power switch at the front panel is turned off.
N O T E
The detector was shipped with default configuration settings. To change these settings see
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......