
144
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
Removing the Covers
When required
For all repairs inside the detector
Tools required
Screwdriver POZI 1 PT3
Parts required
Depends on the work inside and the following procedures
C A U T I O N
There will be a risk of damaging hardware due to overheating when operating the
instrument without covers.
Preparations for this procedure:
•
Turn off the detector.
•
Disconnect the power cable.
•
Remove the flow cell.
•
Remove detector from stack and place it on the
workbench.
1
Press the release buttons and remove the front cover.
Unclip and remove the waste funnel assemly.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......