256
1100 Series FD Reference Manual
7
Introduction to the Fluorescence Detector
Electrical Connections
•
The GPIB connector is used to connect the detector with a computer. The
address and control switch module next to the GPIB connector determines
the GPIB address of your detector. The switches are preset to a default
address (see
on page 291) which is recognized
once the power is switched on.
•
The CAN bus is a serial bus with high speed data transfer. The two
connectors for the CAN bus are used for internal Agilent 1100 Series
module data transfer and synchronization.
•
Two independent analog outputs provide signals for integrators or data
handling systems.
•
The interface board slot is used for external contacts and BCD bottle
number output or LAN connections.
•
The REMOTE connector may be used in combination with other analytical
instruments from Agilent Technologies if you want to use features such as
start, stop, common shut down, prepare, and so forth.
•
With the appropriate software, the RS-232C connector may be used to
control the detector from a computer through a RS-232C connection. This
connector is activated and can be configured with the configuration switch
next to the GPIB connector (see
“Communication Settings for RS-232C
on page 292). See your software documentation for
further information.
Together with a control module the RS-232C may be used to print screens to
a connected printer.
•
The power input socket accepts a line voltage of 100 – 120 V or
220 – 240 V AC ± 10 % with a line frequency of 50 or 60 Hz. Maximum power
consumption is 220 VA. There is no voltage selector on your detector
because the power supply has a wide-ranging capability. There are no
externally accessible fuses, because automatic electronic fuses are
implemented in the power supply. The security lever at the power input
socket prevents removal of the detector cover when line power is still
connected.
Содержание 1100 Series
Страница 1: ...s1 Agilent 1100 Series Fluorescence Detector Reference Manual Agilent G1321A Asset 304025 ...
Страница 12: ...12 1100 Series FD Reference Manual ...
Страница 58: ...60 1100 Series FD Reference Manual 2 First Steps with the Fluorescence Detector ...
Страница 126: ...128 1100 Series FD Reference Manual 4 Troubleshooting and Test Functions ...
Страница 260: ...262 1100 Series FD Reference Manual 7 Introduction to the Fluorescence Detector ...
Страница 324: ...326 1100 Series FD Reference Manual 9 Introduction to the Control Module ...
Страница 340: ...342 1100 Series FD Reference Manual A Safety Information ...
Страница 345: ......