Repairing the Fluorescence Detector
5
1100 Series FD Reference Manual
191
1
Carefully remove the bottom foam piece by sliding it out
towards the rear.
2
Unscrew the power supply at the rear of the module.
3
Press down the power switch light pipe to remove it from
the coupler.
4
Remove the power supply completely.
Re-use the coupler on the new power supply.
The repair of the power supply assembly is comprised of exchanging the complete assembly. No serviceable
parts are inside.
Power
switch light
pipe
Coupler