170
1100 Series FD Reference Manual
5
Repairing the Fluorescence Detector
5
Slide out the condenser assembly and place it in a safe
location.
6
Slide in the condenser assembly completely with the
larger lens diameter towards the cutoff filter.
7
Use a 2.5 mm hexagonal key to fix the screw. Replace the
cutoff filter.
8
Carefully replace the front foam.